RT Journal Article SR Electronic T1 The “Lasso-o” tape: stretchability and observer variability in head circumference measurement JF Archives of Disease in Childhood JO Arch Dis Child FD BMJ Publishing Group Ltd and Royal College of Paediatrics and Child Health SP 820 OP 821 DO 10.1136/adc.2004.063743 VO 90 IS 8 A1 J L Bartram A1 A S Rigby A1 P S Baxter YR 2005 UL http://adc.bmj.com/content/90/8/820.abstract AB Head circumference is an important clinical measurement in children. The stretchability of the Lasso-o tape and the intra- and inter-observer reproducibility of measurements using it were investigated. Old Lasso-o tapes stretch significantly. The intra-class coefficients were 0.999 for intra-observer and 0.979 for inter-observer measurements. Nonetheless in 9% of measurements inter-observer variability was over 1 cm.