PT - JOURNAL ARTICLE AU - P. Goddard AU - I. Keith AU - H. Marcovitch AU - N. R. C. Roberton AU - P. Rolfe AU - J. W. Scopes TI - Use of a continuously recording intravascular oxygen electrode in the newborn AID - 10.1136/adc.49.11.853 DP - 1974 Nov 01 TA - Archives of Disease in Childhood PG - 853--860 VI - 49 IP - 11 4099 - http://adc.bmj.com/content/49/11/853.short 4100 - http://adc.bmj.com/content/49/11/853.full SO - Arch Dis Child1974 Nov 01; 49 AB - The use of a continuously recording intravascular Po2 electrode is described. The electrode, based on the Galvanic principle, is built into the tip of a French gauge (FG) 5 umbilical arterial catheter. Its use has considerably improved the management of oxygen therapy in the seriously ill premature newborn infant and has done so with at least a 50% reduction in the amount of blood needed to be removed for gas analysis.