Responses

Oximetry-detected pulsus paradoxus predicts for severity in paediatric asthma
Free
Compose Response

Plain text

  • No HTML tags allowed.
  • Web page addresses and e-mail addresses turn into links automatically.
  • Lines and paragraphs break automatically.
Author Information
First or given name, e.g. 'Peter'.
Your last, or family, name, e.g. 'MacMoody'.
Your email address, e.g. higgs-boson@gmail.com
Your role and/or occupation, e.g. 'Orthopedic Surgeon'.
Your organization or institution (if applicable), e.g. 'Royal Free Hospital'.
Statement of Competing Interests

PLEASE NOTE:

  • Responses are moderated before posting and publication is at the absolute discretion of BMJ, however they are not peer-reviewed
  • Once published, you will not have the right to remove or edit your response. Removal or editing of responses is at BMJ's absolute discretion
  • If patients could recognise themselves, or anyone else could recognise a patient from your description, please obtain the patient's written consent to publication and send them to the editorial office before submitting your response [Patient consent forms]
  • By submitting this response you are agreeing to our full [Response terms and requirements]

Vertical Tabs

Other responses

Jump to comment:

  • Published on:
    Oximetry-detected pulsus paradoxus predicts for severity in paediatric asthma
    • Paul Seddon, Respiratory Paediatrician Royal Alexandra Children's Hospital, Brighton
    • Other Contributors:
      • David Wertheim, Professor, Computer Science

    We read with interest the paper by Krishnan et al.1 and agree that "pulsus paradoxus" (PP) of the oximetry plethysmogram (pleth) may be useful in assessing severity of acute asthma exacerbations in children. The visual assessment they propose is one of a number of approaches which have been used, with variable success in predicting clinical outcomes2,3,4.

    Rather than pulse amplitude variation associated with respiration, figure 2 in the paper1 shows predominately baseline undulation, at a rate of about 1/5 to 1/6 of pulse rate; no time base nor simultaneous respiratory waveform is included. Can the authors thus be sure that the variation is due to respiration, and if it is, could the baseline variations in fact be associated with respiratory-related changes in peripheral blood volume? As the visual pleth display is dependent on processing by the pulse oximeter, it would be helpful to know more about the oximeter used.

    We have monitored respiration using Respiratory Inductance Plethysmography (RIP) bands simultaneously with oximetry pleth in children with acute wheezing illness using a SOMNOscreen plus recorder (SOMNOmedics GmbH, Germany) with Nonin oximeter module (Nonin Medical Inc., USA). We developed software in MATLAB (The MathWorks Inc., USA) for quantifying pleth pulse amplitude to assess pulsus paradoxus analogous variation. Consistent with Krishnan et al., we found that chil...

    Show More