Abstract
Objective
To compare a system that continuously monitors cardiac output by the Fick principle with measurements by the thermodilution technique in pediatric patients.
Design
Prospective direct comparison of the above two techniques.
Setting
Pediatric intensive care unit of a university hospital.
Patients
25 infants and children, aged 1 week to 17 years (median 10 months), who had undergone open heart surgery were studied. Only patients without an endotracheal tube leak and without a residual shunt were included.
Methods
The system based on the Fick principle uses measurements of oxygen consumption taken by a metabolic monitor and of arterial and mixed venous oxygen saturation taken by pulse- and fiberoptic oximetry to calculate cardiac output every 20 s.
Interventions
In every patient one pair of measurements was taken. Continuous Fick and thermodilution cardiac output measurements were performed simultaneously, with the examiners remaining ignorant of the results of the other method.
Results
Cardiac output measurements ranged from 0.21 to 4.55 l/min. A good correlation coefficient was found:r 2=0.98;P<0.001; SEE=0.14 l/min. The bias is absolute values and in percent of average cardiac output was −0.05 l/min or −4.4% with a precision of 0.32 l/ min or 21.3% at 2 SD, respectively. The difference was most marked in a neonate with low cardiac output.
Conclusion
Continuous measurement of cardiac output by the Fick principle offers a convenient method for the hemodynamic monitoring of unstable infants and children.
Article PDF
Similar content being viewed by others
References
Shoemaker WC, Kram HB, Appel PL (1990) Therapy of shock based on pathophysiology, monitoring, and outcome prediction. Crit Care Med 18: S19-S25
Shoemaker WC, Appel PL, Kram HB (1988) Tissue oxygen debt as a determinant of lethal and nonlethal post-operative organ failure. Crit Care Med 16: 1117–1120
Fanconi S, Doherty P, Edmonds J (1985) Pulse oximetry in pediatric intensive care: comparison with measured saturations and transcutaneous oxygen tension. J Pediatr 107: 362–367
Schranz D, Schmitt S, Oelert H, Schmid F, Huth R, Zimmer B, Schuind A, Vogel K, Stopfkuchen H, Jungst BK (1989) Continuous monitoring of mixed venous oxygen saturation in infants after cardiac surgery. Intensive Care Med 15: 228–232
Weyland W, Weyland A, Fritz U, Redecker K, Ensink F-B, Braun U (1994) A new paediatric metabolic monitor. Intensive Care med 20: 51–57
Takala J, Keinänen O, Väsänen RN, Kari A (1989) Measurement of gas exchange in intensive are: laboratory and clinical validation of a new device. Crit Care Med 17: 1041–1047
Carpenter JP, Nair S, Staw I (1985) Cardiac output determination: thermodilution versus a new computerized Fick method. Crit Care Med 13: 576–579
Keinanen O, Takala J, Kari A (1992) Continuous measurement of cardiac output by the Fick principle: clinical validation in intensive care. Crit Care Med 20: 360–365
Hankeln KB, Gronemeyer R, Held A, Bohmert F (1991) Use of continuous noninvasive measurement of oxygen consumption in patients with adult respiratory distress syndrome following shock of various etiologies. Crit Care Med 19: 642–649
Nakanishi N, Yoshioka T, Okano Y, Nishimura T (1993) Continuous Fick cardiac output measurement during exercise by monitoring of mixed venous oxygen saturation and oxygen uptake [see comments]. Chest 104: 419–426
Chwals WJ, Lally KP, Woolley MM (1992) Indirect calorimetry in mechanically ventilated infants and children: measurement accuracy with absence of audible air leak. Crit Care Med 20: 768–770
Rasanen J (1992) Continuous breathing circuit flow and tracheal tube cuff leak: sources of error during pediatric indirect calorimetry. Crit Care Med 20: 1335–1340
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I: 307–310
Stetz CW, Miller RG, Kelly GE, Raffin TA (1982) Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 126: 1001–1004
Davies GG, Jebson PJ, Glasgow BM, Hess DR (1986) Continuous Fick cardiac output compared to thermodilution cardiac output. Crit Care Med 14: 881–885
Chang AC, Kulik TJ, Hickey PR, Wessel DL (1993) Real-time gas-exchange measurement of oxygen consumption in neonates and infants after cardiac surgery. Crit Care Med 21: 1369–1375
Hillis LD, Firth BG, Winniford MD (1985) Analysis of factors affecting the variability of Fick versus indicator dilution measurements of cardiac output. Am J Cardiol 56: 764–768
Baxley WA, Cavender JB, Knoblock J (1993) Continuous cardiac output monitoring by the Fick method. Cathet Cardiovasc Diagn 28: 89–92
Feustel PJ, Perkins RJ, Oppenlander JE, Stratton HH, Cohen IL (1994) Feasibility of continuous oxygen delivery and cardiac output measurement by application of the Fick principle. Am J Respir Crit Care Med 149: 751–758
Davies G, Hess D, Jebson P (1987) Continuous Fick cardiac output compared to continuous pulmonary artery electromagnetic flow measurement in pigs. Anesthesiology 66: 805–809
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Wippermann, C.F., Huth, R.G., Schmidt, F.X. et al. Continuous measurement of cardiac output by the Fick principle in infants and children: Comparison with the thermodilution method. Intensive Care Med 22, 467–471 (1996). https://doi.org/10.1007/BF01712169
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01712169