Incubator temperature control: effects on the very low birthweight infant.
We studied temperature stability in 22 infants of birthweight less than 1500 g in the first four days of life. Infants were nursed in incubators using either air mode control or skin temperature servo control. Data were collected continuously using a computer linked monitoring system. Skin temperature control resulted in a less stable thermal environment than air mode control. Increased thermal stability in the incubator on air mode control may well be beneficial, particularly to sick, very low birthweight infants.
This article has been cited by other articles:
-
Lyon, A J, Pikaar, M E, Badger, P, McIntosh, N
(1997). Temperature control in very low birthweight infants during first five days of life. Arch. Dis. Child. Fetal Neonatal Ed.
76: 47F-50
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.